Verification test requirements
Be sure that you perform the verification tests:
Under the proper environmental conditions.
After the specified warm-up period.
Using the correct line voltage.
Using the proper calibration equipment.
Using the specified reading limits.
Environmental conditions
Conduct your performance verification procedures in a test environment that has:
An ambient temperature of 18-28°C (65-82°F).
A relative humidity of less than 80% unless otherwise noted.
Line power
The Model 2000 Multimeter requires a line voltage of 100V/120V/220V/240V, ±10% and a
line frequency of 45Hz to 66Hz, or 360Hz to 440Hz.
Warm-up period
Allow the Model 2000 Multimeter to warm up for at least one hour before conducting the ver-
ification procedures.
If the instrument has been subjected to temperature extremes (those outside the ranges stated
above), allow additional time for the instrument’s internal temperature to stabilize. Typically,
allow one extra hour to stabilize a unit that is 10°C (18°F) outside the specified temperature
range.
Also, allow the test equipment to warm up for the minimum time specified by the manufac-
turer.
Performance Verification 1-3

Table of Contents

Other manuals for Keithley 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2000 and is the answer not in the manual?

Keithley 2000 Specifications

General IconGeneral
Digits6.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range100 Ω to 100 MΩ
ConnectivityIEEE-488 (GPIB)
Display TypeVFD

Related product manuals