EasyManuals Logo

Keithley 2000 Calibration Manual

Keithley 2000
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #25 background imageLoading...
Page #25 background image
Environmental conditions
Conduct the calibration procedures in a location that has:
An ambient temperature of 23°C ±5°C
A relative humidity of less than 80% unless otherwise noted
Warm-up period
Allow the Model 2000 Multimeter to warm up for at least one hour before performing cali-
bration.
If the instrument has been subjected to temperature extremes (those outside the ranges stated
in the above section) allow extra time for the instrument’s internal temperature to stabilize. Typ-
ically, allow one extra hour to stabilize a unit that is 10°C (18°F) outside the specified tempera-
ture range.
Also, allow the test equipment to warm up for the minimum time specified by the manufac-
turer.
Line power
The Model 2000 Multimeter requires a line voltage of 100V/120V/220V/240V, ±10% and a
line frequency of 45Hz to 66Hz or 360Hz to 440Hz.
Calibration 2-3

Table of Contents

Other manuals for Keithley 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2000 and is the answer not in the manual?

Keithley 2000 Specifications

General IconGeneral
Digits6.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range100 Ω to 100 MΩ
ConnectivityIEEE-488 (GPIB)
Display TypeVFD

Related product manuals