Front panel calibration
Follow the steps in the following paragraphs for comprehensive, DC only, and AC only cali-
bration procedures.
The procedures for front panel calibration include:
Preparing the Model 2000 for calibration
Front panel short and open calibration
DC voltage calibration
Resistance calibration
DC current calibration
AC voltage calibration
AC current calibration
Setting calibration dates
Preparing the Model 2000 for calibration
1. Turn on the Model 2000, and allow it to warm up for at least one hour before performing
the calibration procedure.
2. Select the DCV function and choose SLOW as the RATE (integration time = 10 PLC).
3. Start the calibration process as follows:
A. Access the calibration menu by pressing SHIFT then CAL.
B. Use the up and down range keys to scroll through the available calibration menu
items until the unit displays RUN, then press ENTER.
C. At the prompt, enter the calibration code. (The default code is 002000.) Use the left
and right arrow keys to move among the digits; use the up range key to increment
numbers, and press the down range key to specify alphabetic letters. Confirm the
code by pressing ENTER.
D. Choose N at the prompt to proceed without changing the code, then press ENTER.
4. Choose which of the the calibration tests summarized in Table 2-2 you want to run at the
CAL: RUN prompt. Use the up and down range keys to scroll through the options; select
your choice by pressing ENTER.
Table 2-2
Comprehensive calibration procedures
Procedure Menu choice Procedures
Full calibration
DCV, DCI, and ohms
ACV and ACI
ALL
DC
AC
All comprehensive calibration steps.
DC voltage, DC current, and resistance calibration.
AC voltage and AC current calibration.
Calibration 2-7

Table of Contents

Other manuals for Keithley 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2000 and is the answer not in the manual?

Keithley 2000 Specifications

General IconGeneral
Digits6.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range100 Ω to 100 MΩ
ConnectivityIEEE-488 (GPIB)
Display TypeVFD

Related product manuals