DC calibration commands
The :DC commands perform comprehensive (user) calibration of the DCV, DCI, and ohms
functions. Table 3-2 summarizes these comprehensive calibration commands along with param-
eter limits.
Table 3-2
DC calibration commands
Command Description
Parameter
limits
:CALibration
:PROTected
:DC
:STEP1
:STEP2
:STEP3 <NRf>
:STEP4 <NRf>
:STEP5 <NRf>
:STEP6 <NRf>
:STEP7 <NRf>
:STEP8 <NRf>
:STEP9 <NRf>
:STEP10 <NRf>
:STEP11 <NRf>
:STEP12 <NRf>
Front terminal short circuit.
Open circuit.
10V DC calibration step.
-10V DC calibration step.
100V DC calibration step.
1k
4-wire calibration step.
10k
4-wire calibration step.
100k
4-wire calibration step.
1M
4-wire calibration step.
10mA DC calibration step.
100mA DC calibration step.
1A DC calibration step.
9 to 11
-9 to -11
90 to 110
900 to 1.1E3
9E3 to 11E3
90E3 to 110E3
900E3 to 1.1E6
9E-3 to 11E-3
90E-3 to 110E-3
0.9 to 1.1
3-10 Calibration Command Reference

Table of Contents

Other manuals for Keithley 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2000 and is the answer not in the manual?

Keithley 2000 Specifications

General IconGeneral
Digits6.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range100 Ω to 100 MΩ
ConnectivityIEEE-488 (GPIB)
Display TypeVFD

Related product manuals