:STEP1
(:CALibration:PROTected:DC:STEP1)
Purpose
To perform front terminal short-circuit calibration.
Format
:cal:prot:dc:step1
Parameter
None
Description :STEP1 performs the short-circuit calibration step in the comprehensive cal-
ibration procedure. Connect a low-thermal short (Model 8610) to the front
panel input jacks before sending this command.
Example :CAL:PROT:DC:STEP1 Perform short-circuit calibration.
:STEP2 (:CALibration:PROTected:DC:STEP2)
Purpose To perform front terminal open-circuit calibration.
Format :cal:prot:dc:step2
Parameter None
Description :STEP2 performs the open-circuit calibration step in the comprehensive cal-
ibration procedure. Disconnect all cables and accessories from the input
jacks before sending this command.
Example :CAL:PROT:DC:STEP2 Perform open circuit calibration.
:STEP3 (:CALibration:PROTected:DC:STEP3)
Purpose To program the +10V comprehensive calibration step.
Format :cal:prot:dc:step3 <Cal_voltage>
Parameter <Cal_voltage> = 9 to 11 [V]
Description :STEP3 programs the +10V DC comprehensive calibration step. The allow-
able range of the calibration voltage parameter is from 9 to 11, but 10 is rec-
ommended for best results.
Example :CAL:PROT:DC:STEP3 10 Program 10V step.
Calibration Command Reference 3-11

Table of Contents

Other manuals for Keithley 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2000 and is the answer not in the manual?

Keithley 2000 Specifications

General IconGeneral
Digits6.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range100 Ω to 100 MΩ
ConnectivityIEEE-488 (GPIB)
Display TypeVFD

Related product manuals