Introduction
Use the procedures in this section to calibrate the Model 2000. Calibration procedures in-
clude:
Comprehensive calibration:
Calibrate DC and AC voltages, DC and AC currents, and
resistance values.
Manufacturing calibration:
Usually only performed at the factory.
WARNING This information in this section is intended only for qualified service person-
nel. Do not attempt these procedures unless you are qualified to do so.
All the procedures require accurate calibration equipment to supply precise DC and AC volt-
ages, DC and AC currents, and resistance values. Comprehensive AC and DC calibration can be
performed any time by a technician either from the front panel, or by using the SCPI commands
sent either over the IEEE-488 bus or the RS-232 link.
NOTE
Manufacturing calibration is required in the field only if the Model 2000 has been re-
paired.
This section includes the following information:
Environmental conditions:
Explains the type of environment needed for calibration.
Calibration considerations:
Summarizes test conditions to observe when performing cali-
bration.
Calibration code:
Explains how to enter the calibration code to unlock Model 2000 calibra-
tion.
Comprehensive calibration:
Summarizes the calibration cycle and also lists recommended
comprehensive calibration equipment.
Front panel calibration:
Provides the calibration procedures using the front panel.
SCPI command calibration:
Provides the detailed procedures for calibrating the Model
2000 using SCPI commands.
Manufacturing calibration:
Explains the manufacturing calibration procedure using both
the front panel and SCPI commands.
2-2 Calibration

Table of Contents

Other manuals for Keithley 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2000 and is the answer not in the manual?

Keithley 2000 Specifications

General IconGeneral
Digits6.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range100 Ω to 100 MΩ
ConnectivityIEEE-488 (GPIB)
Display TypeVFD