EasyManuals Logo

Keithley 2000 Calibration Manual

Keithley 2000
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #15 background imageLoading...
Page #15 background image
Performing the verification test procedures
Test summary
Verification test procedures include:
DC volts
AC volts
DC current
AC current
Resistance
If the Model 2000 is not within specifications and not under warranty, see the calibration pro-
cedures in Section 2.
Test considerations
When performing the verification procedures:
Be sure to restore factory defaults as outlined above.
Make sure that the equipment is properly warmed up and connected to the front panel
input jacks. Also make sure that the front panel input jacks are selected with the INPUTS
switch.
Do not use autoranging for any verification tests because autorange hysteresis may cause
the Model 2000 to be on an incorrect range. For each test signal, you must manually set
the correct range for the Model 2000 using the range keys.
Make sure the calibrator is in operate before you verify each measurement.
Always let the source signal settle before taking a reading.
Do not connect test equipment to the Model 2000 through a scanner or other switching
equipment.
WARNING The maximum common-mode voltage (voltage between INPUT LO and chas-
sis ground is 500 V peak. Exceeding this value may cause a breakdown in
insulation, creating a shock hazard. Some of the procedures in this section
may expose you to dangerous voltages. Use standard safety precautions
when such dangerous voltages are encountered to avoid personal injury
caused by electric shock.
1-6 Performance Verification

Table of Contents

Other manuals for Keithley 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2000 and is the answer not in the manual?

Keithley 2000 Specifications

General IconGeneral
Digits6.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range100 Ω to 100 MΩ
ConnectivityIEEE-488 (GPIB)
Display TypeVFD

Related product manuals