:NDUE
:CALibration:PROTected:NDUE)
Purpose
To send the next calibration due date to the instrument.
Format
:cal:prot:ndue <year>, <month>, <day>
Parameters
<year> = 1994 to 2093
<month> = 1 to 12
<day> = 1 to 31
Query format
:cal:prot:ndue?
Response
<year>, <month>, <day>
Description
The :NDUE command allows you to store the date when calibration is next
due in instrument memory. You can read back the next due date from the in-
strument over the bus by using the :NDUE? query or the front panel CAL
menu.
Note
The next due date parameters must be delimited by commas.
Examples
:CAL:PROT:NDUE 1995,12,16 Send due date (12/16/95).
:CAL:PROT:NDUE? Request due date.
Calibration Command Reference 3-9