:CALibration
:PROTected
:AC
:STEP1
:STEP2
:STEP3
:STEP4
:STEP5
:STEP6
:STEP7
:STEP8
:STEP9
:STEP10
:STEP11
:STEP12
:STEP13
:STEP14
:STEP15
Calibration root command.
AC cal steps.
10mV AC at 1kHz step.
100mV AC at 1kHz step.
100mV AC at 50kHz step.
1V AC at 1 kHz step.
1V AC at 50kHz step.
10V AC at 1kHz step.
10V AC at 50kHz step.
100V AC at 1kHz step.
100V AC at 50kHz step.
700V AC at 1kHz step.
100mA AC at 1kHz step.
1A AC at 1kHz step.
2A AC at 1kHz step.
1V AC at 3Hz step.
1
1V AC at 1kHz step.
1
NOTES:
1. DC:STEP0, AC:STEP14, and AC:STEP15 are one-time factory calibration points and are only valid in
manufacturing calibration mode.
2. Upper case letters indicated short form of each command. For example, instead of sending ":CALibra-
tion:PROTected:INITiate," you can send":CAL:PROT:INIT."
Table 3-1 (cont.)
Calibration command summary
Command Description
Calibration Command Reference 3-5