:CALibration
:PROTected
:AC
:STEP1
:STEP2
:STEP3
:STEP4
:STEP5
:STEP6
:STEP7
:STEP8
:STEP9
:STEP10
:STEP11
:STEP12
:STEP13
:STEP14
:STEP15
Calibration root command.
AC cal steps.
10mV AC at 1kHz step.
100mV AC at 1kHz step.
100mV AC at 50kHz step.
1V AC at 1 kHz step.
1V AC at 50kHz step.
10V AC at 1kHz step.
10V AC at 50kHz step.
100V AC at 1kHz step.
100V AC at 50kHz step.
700V AC at 1kHz step.
100mA AC at 1kHz step.
1A AC at 1kHz step.
2A AC at 1kHz step.
1V AC at 3Hz step.
1
1V AC at 1kHz step.
1
NOTES:
1. DC:STEP0, AC:STEP14, and AC:STEP15 are one-time factory calibration points and are only valid in
manufacturing calibration mode.
2. Upper case letters indicated short form of each command. For example, instead of sending ":CALibra-
tion:PROTected:INITiate," you can send":CAL:PROT:INIT."
Table 3-1 (cont.)
Calibration command summary
Command Description
Calibration Command Reference 3-5

Table of Contents

Other manuals for Keithley 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2000 and is the answer not in the manual?

Keithley 2000 Specifications

General IconGeneral
Digits6.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range100 Ω to 100 MΩ
ConnectivityIEEE-488 (GPIB)
Display TypeVFD