2-22 Calibration
Resetting the calibration code
If you lose the calibration code, you can unlock calibration by temporarily shorting together
the CAL pads, which are located on the digital board, while turning on the power. Since the dig-
ital board is located under the analog board, it may be necessary to use extender wires or to
remove the analog board to gain access.
Viewing calibration date and count
Viewing date and count from the front panel
Follow the steps below to view the Model 2303 or 2303-PJ calibration date and count from
the front panel:
1. Press the MENU key, then choose CALIBRATE UNIT, and press ENTER. The instru-
ment will display the last date calibrated:
CALIBRATE UNIT
LAST ON 02/01/97
2. Press the up arrow key. The instrument will display the number of times it was calibrated:
CALIBRATE UNIT
TIMES= 01
3. Press MENU to return to the menu structure.
Acquiring date and count by remote
Use the :DATE? and :COUNT? queries to determine the Model 2303, 2303B, or 2303-PJ cal-
ibration date and count respectively. See Miscellaneous commands in Appendix B for more
details.
深圳德标仪器 135-1095-0799

Table of Contents

Other manuals for Keithley 2303

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2303 and is the answer not in the manual?

Keithley 2303 Specifications

General IconGeneral
BrandKeithley
Model2303
CategoryPower Supply
LanguageEnglish

Related product manuals