EasyManuals Logo

Keithley 2303 Service Manual

Keithley 2303
104 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #89 background imageLoading...
Page #89 background image
B-8 Calibration Reference
Generating an SRQ on calibration complete
An IEEE-488 bus SRQ (service request) can be used to detect operation complete instead of
repeatedly polling the Model 2303. To use this method, send both *ESE 1 and *SRE 32 to the
instrument, then include the *OPC command at the end of each calibration command line, as
covered above. Clear the SRQ by querying the ESR (using the *ESR? query) to clear OPC sta-
tus, then request the status byte with the *STB? query to clear the SRQ.
Refer to your controller's documentation for information on detecting and servicing SRQs.