Model 2750 Multimeter/Switch System Service Manual Performance Verification 1-33
7700 Verification
RTD temperature
1. Connect the precision decade resistance box (listed in Table 1-1) to the Model
7700 CH1 and CH11 H and L terminals using four-wire connections. (See
Figure 1-14 for similar connecting scheme.)
2. Install the Model 7700 in Slot 1 of the Model 2750, then turn on the power and
allow the unit to warm up for two hours before proceeding. Be sure the front panel
INPUTS switch is set to the REAR position.
3. Select the temperature function by pressing the TEMP key. Close Channel 1 by
pressing the CLOSE key and keying in 101.
4. Configure the Model 2750 temperature function for °C units and RTD temperature
sensor (=0.00385) as follows:
a. Press SHIFT then SENSOR, and note the unit displays the temperature units:
UNITS: C.
b. Press ENTER, and note the unit displays the sensor type: SENS: TCOUPLE.
c. Using the cursor and range keys, set the display as follows: SENS: 4W-RTD.
d. Press ENTER, and note the unit displays: TYPE: PT100.
e. Using the cursor and range keys, set the unit for the following display: TYPE:
PT385.
f. Press ENTER to complete the temperature configuration process.
5. Set the decade resistance box to each of the values shown in Table 1-17, and verify
that the temperature readings are within the required limits. Open Channel 1 when
finished.
Tab le 1- 17
Plug-in module four-wire RTD temperature verification reading limits
Applied Resistance* Reading Limits (1 year, 18°C to 28°C)
22.80
100.00
313.59
-190.06 to -189.94°C
-0.06 to +0.06°C
599.94 to 600.06°C
*Based on = 0.00385. See text.

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A

Related product manuals