Table 3-11. Device-Dependent Command Summary
-
/lode
Command
Description
Paragrapl
xecute
X Execute other device-dependent commands.
3.10.1
:unction
FO Volts
3.102
Fl
Amps
F2 Ohms
F3 Coulombs
F4
External Feedback
F5 V/I Ohms
lange
: External V/I
3.10.3
Volts Amps Ohms Coulombs Feedback Ohms
RO Auto
Auto Auto Auto Auto
Auto
Rl 200mV
2 PA
2 kR 2OOpC 200mV 2OOTQ
R2
2 V 20 pA 20 kR 2nC 2V 20Tfl
R3 20 V 200 pA 200 k0
20 nc 20 v 2TR
R4
200 v
2 nA 2MQ 20 nC 20 v
200GLl
R5 200 V 20 nA 20M0 20 nC 20 v
20G$l
R6 200 V 200 nA 200MQ
20 nC 20 v 2G0
R7
200 V 2 /LA 2GO
20 nC 20 v
200MR
R6
200 V 20/~A 20GS2
20 nC 20 v
20MR
R9
200 V 200 PA 200G0
20 nC 20 v
2MR
RlO 200 v 2mA 200GSI
20 nC
20 v
200 kR
Rll 200 V 20mA 200GQ
20 nC
20 v
200 kQ
R12 Cancel autoranging for all functions
:ero Check
co
Zero Check Off
3.10.4
Cl
Zero Check On
:ero correct
zo Zero Correct Disabled
3.10.4
Zl Zero Correct Enabled
laseline Suppression
NO Suppression Disabled
3.10.5
Nl Suppression Enabled
display Mode
DO Electrometer
3.10.6
Dl
Voltage Source
leading Mode
BO Electrometer
3.10.7
El
Buffer Reading
82
Maximum Reading
83 Minimum Reading
84
Voltage Source
)ata Store
00 Conversion rate
3.10.8
01
One Reading Per Second
a2 One Reading Every 10 Seconds
Q3 One Reading Per Minute
Q4 One Reading Every 10 Minutes
Q5
One Reading Per Hour
Q6 Trigger Mode
Q7 Disabled
loltage Source
V+ nnn.nn or Voltage Source Value: - 102.35V to + 102.4V, 50mV
3.10.9
increments
falue
V+n.nnnnE+n
roltage Source
00 Source Output Dff (OV)
3.10.10
)perate
01 Source Output On (Programmed Value)
Calibration Value
A+ nnn.nn or Calibrate Function and Range
3.10.11
An.nnnE+n
: .^.^ P”l:b.-^ri^^
I4 Ph.-,. P”liL.....i.... P^..^.^^.^ i.. hl\,D Ah”
2 ‘1n ‘I?
3-19