EasyManua.ls Logo

Keithley 6517 - Figure 2-3 Connections for 20 Pa-2Μa Range Calibration

Keithley 6517
119 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration
2-9
4. Connect the Model 5156 Electrometer Calibration Stan-
dard and the DC voltage calibrator to the Model 6517A
INPUT jack, as shown in Figure 2-3. Initially, make
connections to the 100G
resistance in the standards
box.
5. After the zero current calibration step, the instrument
will prompt you as follows:
CONNECT 2V to 100G
ENTER to continue; EXIT to abort
6. Set the DC voltage calibrator to exactly 1.900000V DC.
Make sure the 100G
resistor in the calibration stan-
dard is connected, then press the Model 6517A ENTER
key. The instrument will prompt for the exact calibration
value:
1.9000000 V
Use
,
, , ,ENTER,EXIT or INFO
NOTE
For all calibration steps that involve the
Model 5156, you can set calibration values
in either one of two ways: (1) set the cali-
brator output to agree with the displayed
value, or (2) adjust the display to agree
with the calibrator value.
Figure 2-3
Connections for 20pA–2µA range calibration
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
INPUT
250V PEAK
!
LINE RATING
50-60HZ
50VA MAX
AC ONLY
LINE FUSE
SLOWBLOW
1/2A 90-125V
1/4A 180-250V
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
DIGITAL
I/O
TRIG LINK
115V
LO
HI
RS232
!
MADE
IN
U.S.A.
PREAMP OUT
250 PEAK
COMMON 2V ANALOG
OUTPUT
TEMP
TYPE K
HUMIDITY
OPTION SLOT
V SOURCE
EXT TRIG
IN
MTR COMP
OUT
INTERLOCK
SELECTED
LINE VOLTAGE
90-110V
105-125V
180-220V
210-250V
1010V
PEAK
750V
PEAK
Low-noise
Coax Cable
Triax Cable
Model 6517 Electrometer
Connect Cable
Shield to
Output LO
BNC-to-dual
Banana Plug
Adapter
DC Voltage Calibrator
Note: Connect Calibrator to
Appropriate Resistor.
Link Chassis and Shield
(See Text)
100G
1nF
100nF
10G
1G
100M
Model 5156 Electrometer Calibration Standard
OUTPUT
!
7. Adjust the calibrator voltage to agree with the exact dis-
play value, then press the ENTER key. During this step,
the instrument will display the following:
Performing 20 pA cal
8. Next, the instrument will prompt you as follows:
CONNECT -2V to 100G
ENTER to continue; EXIT to abort
9. Press the Model 6517A ENTER key. The instrument
will prompt for the exact calibration value:
-1.9000000 V
Use , , , ,ENTER,EXIT or INFO
10. Set the calibrator output to the display value, then press
the ENTER key. The unit will display the following dur-
ing this calibration step:
Performing -20 pA Calibration
11. Repeat steps 5 through 10 for the 200pA through 2µA
ranges using the voltages and resistance standards sum-
marized in Table 2-4.
12. Disconnect the calibration standard and voltage calibra-
tor from the instrument, and connect the DC current cal-
ibrator directly to the Model 6517A INPUT jack (see
Figure 2-4).
13. At this point, the Model 6517A will display the
following:
CONNECT 19.00000 µA
ENTER to continue; EXIT to abort

Table of Contents

Related product manuals