Calibration Command Reference
3-11
3.4.3 Amps calibration commands
Purpose To calibrate amps function ranges.
Format See Tables 3-3 and 3-4.
Parameter See Tables 3-3 and 3-4.
Table 3-3
Amps function calibration commands and parameters (all ranges, without calibration standard)
Range Command format* <Cal_current> parameter limits (A)
20pA :cal:prot:azero20p
:cal:prot:a20p <Cal_current>
:cal:prot:an20p <Cal_current>
None
9.5E-12 to 20.5E-12
-9.5E-12 to -20.5E-12
200pA :cal:prot:azero200p
:cal:prot:a200p <Cal_current>
:cal:prot:an200p <Cal_current>
None
95E-12 to 205E-12
-95E-12 to -205E-12
2nA :cal:prot:azero2n
:cal:prot:a2n <Cal_current>
:cal:prot:an2n <Cal_current>
None
0.95E-9 to 2.05E-9
-0.95E-9 to -2.05E-9
20nA :cal:prot:azero20n
:cal:prot:a20n <Cal_current>
:cal:prot:an20n <Cal_current>
None
9.5E-9 to 20.5E-9
-9.5E-9 to -20.5E-9
200nA :cal:prot:azero200n
:cal:prot:a200n <Cal_current>
:cal:prot:an200n <Cal_current>
None
95E-9 to 205E-9
-95E-9 to -205E-9
2µA :cal:prot:azero2u
:cal:prot:a2u <Cal_current>
:cal:prot:an2u <Cal_current>
None
0.95E-6 to 2.05E-6
-0.95E-6 to -2.05E-6
20µA :cal:prot:azero20u
:cal:prot:a20u <Cal_current>
:cal:prot:an20u <Cal_current>
None
9.5E-6 to 20.5E-6
-9.5E-6 to -20.5E-6
200µA :cal:prot:azero200u
:cal:prot:a200u <Cal_current>
:cal:prot:an200u <Cal_current>
None
95E-6 to 205E-6
-95E-6 to -205E-6
2mA :cal:prot:azero2m
:cal:prot:a2m <Cal_current>
:cal:prot:an2m <Cal_current>
None
0.95E-3 to 2.05E-3
-0.95E-3 to -2.05E-3
20mA :cal:prot:azero20m
:cal:prot:a20m <Cal_current>
:cal:prot:an20m <Cal_current>
None
9.5E-3 to 20.5E-3
-9.5E-3 to -20.5E-3
* Command short form shown.