Calibration Command Reference
3-16
3.6 Temperature calibration commands
Purpose To calibrate the temperature function.
Format :cal:prot:tzero 0mV (0°C) cal point
:cal:prot:t100 4.095mV (100°C) cal point
Parameter None
Description The :T commands calibrate the Model 6517A temperature function. The only two temperature
calibration points correspond to 0mV input and 4.095mV input respectively. A suitable refer-
ence junction compensated type K thermocouple calibrator is required for calibration.
Example :CAL:PROT:TZERO Cal 0mV temperature point.
3.7 Humidity calibration commands
Purpose To calibrate the humidity function.
Format :cal:prot:humzero 0V cal point
:cal:prot:hum05 0.5V cal point
:cal:prot:hum1 1V cal point
Parameter None
Description The :HUM commands calibrate the Model 6517A humidity function. Three calibration points
are required: 0V, 0.5V, and 1V, which correspond to 0%, 50%, and 100% relative humidity
respectively.
Example :CAL:PROT:HUM05 Cal 50% humidity point.