Calibration Command Reference
3-17
3.8 Calibration errors
3.8.1 Error query commands
Purpose To request calibration errors over the bus.
Format :cal:prot:eerr? Request cal execution error status
:cal:prot:verr? Request voltage function cal errors
:cal:prot:aerr? Request current function cal errors
:cal:prot:cerr? Request charge function cal errors
:cal:prot:terr? Request temperature function cal errors
:cal:prot:herr? Request humidity function cal errors
:cal:prot:serr? Request voltage source cal errors
:cal:prot:ferr? Request factory cal errors
Parameter None
Description The calibration error queries allow you to access the various calibration error messages that
might be generated during calibration. A separate query is provided for each function or cate-
gory. For example, the :cal:prot:verr? query is used to request only those errors that occur while
calibrating the voltage function.
Programming Note See Appendix B for a complete listing of calibration error messages.
Examples :CAL:PROT:AERR? Request amps cal errors.
:CAL:PROT:SERR? Request voltage source cal errors.