EasyManua.ls Logo

Keithley 6517 - Coulombs Calibration Summary; Figure 2-6 Connections for Temperature Calibration; Table

Keithley 6517
119 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration
2-12
Step 6: Temperature Calibration
1. Connect the thermocouple calibrator to the Model
6517A EXT TEMP jack, as shown in Figure 2-6.
2. At the end of the coulombs calibration phase, the instru-
ment will prompt you for the first temperature calibra-
tion point:
CONNECT 0V/0°C
ENTER to continue; EXIT to abort
Table 2-6
Coulombs calibration summary
6517A
range
Calibration
voltage
Standard
1
capacitance
Nominal
2
charge
2nC
2nC
20nC
20nC
200nC
200nC
2µC
2µC
1.900000V
-1.900000V
19.00000V
-19.00000V
1.900000V
-1.900000V
19.00000V
-19.00000V
1nF
1nF
1nF
1nF
100nF
100nF
100nF
100nF
1.9nC
-1.9nC
19nC
-19nC
190nC
-190nC
1.9µC
-1.9µC
1
Nominal capacitance standard shown. Refer to calibration data for
actual value.
2
Charge calculated from: Q = CV, where C is capacitance standard
value, and V is the calibrator voltage. When using the Model 5156
Electrometer Calibration Standard, the Model 6517A automatically
calculates the charge from the actual capacitance value and the applied
calibrator voltage.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
INPUT
250V PEAK
!
LINE RATING
50-60HZ
50VA MAX
AC ONLY
LINE FUSE
SLOWBLOW
1/2A 90-125V
1/4A 180-250V
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
DIGITAL
I/O
TRIG LINK
115V
LO
HI
RS232
!
MADE
IN
U.S.A.
PREAMP OUT
250 PEAK
COMMON 2V ANALOG
OUTPUT
TEMP
TYPE K
HUMIDITY
OPTION SLOT
V SOURCE
EXT TRIG
IN
MTR COMP
OUT
INTERLOCK
SELECTED
LINE VOLTAGE
90-110V
105-125V
180-220V
210-250V
1010V
PEAK
750V
PEAK
EXT Temp
Jack
Type K
Thermocouple Wire
Model 6517 Electrometer
Type K
Thermocouple Simulator/Calibrator
Figure 2-6
Connections for temperature calibration
3. Set the thermocouple calibrator output to 0°C (0mV),
then press the Model 6517A ENTER key. During this
step, the instrument will display the following:
Performing 0 V Temp Calibration
4. Next, the instrument will prompt you for the 100°C
(4.095mV) calibration point:
CONNECT 4.095mV/100°C
5. Set the thermocouple calibrator output to 100°C, then
press the Model 6517A ENTER key. During this step,
the instrument will display the following message:
Performing 4.095mV Temp Calibration
Step 7: Voltage Source Calibration
WARNING
Hazardous voltages will be present
when performing the following steps.
Avoid touching terminals while per-
forming these procedures.
1. After temperature calibration has been completed, the
instrument will prompt you to connect the voltmeter to
the voltage source output jacks:
V-SOURCE CALIBRATION
Connect Vsource to voltmeter
2. Select the DCV function and the auto-range mode on the
DMM.

Table of Contents

Related product manuals