EasyManuals Logo

KERN DBS 60-3 User Manual

KERN DBS 60-3
91 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #33 background imageLoading...
Page #33 background image
DBS-BA-e-1311 33
6.2 Adjust temperature
We recommend sometimes to check the temperature value of the device using the
optional temperature calibrating set DBS-A01. The temperature is measured at two
points (100°C and 180°C) and can be adjusted at these two points.
The equipment should cool down first to room temperature after the last heating
period.
Preparation:
Fix the temperature sensor on the
temperature calibration set in
accordance with the illustration.
Remove sample dish and dish holder
from the moisture analyser.
Install the temperature-calibration set
acc. to fig.
Close heated cover.
If applicable, the error message “ol” will
be displayed, but can be ignored in this
case.
On the temperature calibration set
switch-on the digital thermometer by the
ON/OFF button.

Table of Contents

Other manuals for KERN DBS 60-3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN DBS 60-3 and is the answer not in the manual?

KERN DBS 60-3 Specifications

General IconGeneral
BrandKERN
ModelDBS 60-3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals