EasyManuals Logo

KERN DBS 60-3 User Manual

KERN DBS 60-3
91 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #45 background imageLoading...
Page #45 background image
DBS-BA-e-1311 45
9.2 Adjust
9.2.1 Drying modes
9.2.1.1 Drying mode AUTO
(Standard drying/switch-off criterion “M“)
Drying will be carried out at the drying temperature preset by the user.
The drying is finished automatically when the preset weight loss (M ) remains
constant for 30 seconds.
SETTINGS:
Drying mode AUTO
Drying temperature
M
Press the Menu button to access the menu and
the first menu item “PRoGRM” will be displayed.
Press navigation button , then use navigation
buttons to selectAUTO.
Press ENTER to confirm and the currently preset
drying temperature will flash in the display.
Use navigation buttons to select the
required setting,
50°C 200°C (1°C increments) can be selected.
Press ENTER to confirm selection. The currently
preset value for M will be displayed.
Use navigation buttons to select the
required setting, 0.01% 0.1% (0.01%
increments) can be selected.
Press ENTER to save.
Press ESC to return to Moisture analysis mode.
The measuring can be started (see chap. 8.1).
All the settings that have been made control the
drying process.
The switch-off criterion has been fulfilled with a sample whose moisture content is below
0.01%. In this case, we recommend selecting the drying program TIME described below, see
chap. 9.2.1.2

Table of Contents

Other manuals for KERN DBS 60-3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN DBS 60-3 and is the answer not in the manual?

KERN DBS 60-3 Specifications

General IconGeneral
BrandKERN
ModelDBS 60-3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals