9.2.1.2 Drying mode TIME
(Standard drying/switch-off criterion “Time“)
The drying will be carried out at the drying temperature preset by the user and will
end automatically when the preset time has expired.
SETTINGS:
Drying mode TIME
Drying temperature
Drying time
Press the Menu button to access the menu and
the first menu item “PRoGRM” will be displayed.
Press navigation button , then use navigation
buttons to selectTIME”.
Press ENTER to confirm and the currently preset
drying temperature will flash in the display.
Use navigation buttons to select the
required setting,
50°C 200°C (1°C increments) can be selected.
Press ENTER to confirm selection. The currently
set drying time is displayed.
Use navigation buttons to select the
required setting, 1 minute 4 hours (1 minute
increments) or 4 hours - 12 hours (1 hour
increments) can be selected
Press ENTER to save.
Press ESC to return to Moisture analysis mode.
The equipment is ready to measure
(see chap. 8.1). All the settings that have been
made control the drying process.

Table of Contents

Other manuals for KERN DBS 60-3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN DBS 60-3 and is the answer not in the manual?

KERN DBS 60-3 Specifications

General IconGeneral
BrandKERN
ModelDBS 60-3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals