14.1 Error messages
Error message
Explication
Remedy
ERR.001
ERR.002
Hardware error
Switch equipment off and on
again. If the error message
remains inform manufacturer.
ERR.005 Memory error
Switch equipment off and on
again. If the error message
remains inform manufacturer.
ERR.100
Heated cover is open for more
than 1 minute during
measuring.
Press ESC to abort measuring.
ERR.101
ERR.102
Failure „Temperature sensor
Switch equipment off and on
again. If the error message
remains inform manufacturer.
ERR.110
Heated cover not closed
properly
Press ESC to abort measuring.
TIM.oUT
Measuring start 30 minutes
after resetting to zero
Press ESC to abort measuring.
ERR.121
ERR.122
ERR.123
Failure „Heating
Switch equipment off and on
again. If the error message
remains inform manufacturer.
ERR.124 Measuring lasts too long
Check drying time switch-off
criterion or M
ERR.200 Failure „Current supply“
Switch equipment off and on
again. If the error message
remains inform manufacturer.
ERR.201 Internal error
Switch equipment off and on
again. If the error message
remains inform manufacturer.
ERR.202 Failure „Electrical voltage“
Switch equipment off and on
again. If the error message
remains inform manufacturer.

Table of Contents

Other manuals for KERN DBS 60-3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN DBS 60-3 and is the answer not in the manual?

KERN DBS 60-3 Specifications

General IconGeneral
BrandKERN
ModelDBS 60-3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals