90 DBS-BA-e-1311
ERR.C01
ERR.C02
ERR.C04
Adjustment
error
High zero point shift during
adjustment
Press ESC to abort and repeat
adjustment procedure.
There are objects in the sample
dish
Missing sample dish
ERR.oL
ERR.-oL
Overload Check sample dish
CoM.ERR Wrong remote control order. Correct remote control order.
oL
-OL
Overload
Install sample dish correctly
Reduce sample weight
ABORT Process cancelled
Press ESC to return to
Weighing mode
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.