14 Instant help
Fault Possible cause
Display is not lit up.
The display unit is not switched on.
The mains supply connection has been
interrupted (mains cable not plugged in/faulty).
Power supply interrupted.
Fuse has blown
The display does not change
when a sample is being loaded
Sample dish / dish holder is fitted incorrectly.
The weight display changes
constantly / the stability display
does not appear.
Sample dish has contact with wind protection
device or heated cover.
Draught/air movement
Table/floor vibrations
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
Incorrect measuring result
Check adjustment
No resetting to zero before loading the sample
Measurement is taking too
long
Incorrect setting shutoff criterion
Measurement is
not reproducible
Sample is not homogenous
Drying time is too short
Drying temperature too high (e.g. oxidation
sample material, boiling point of sample
exceeded)
Temperature sensor soiled or defective
Drying does not start
Heated cover open
The mains supply connection has been
interrupted (mains cable not plugged in/faulty).

Table of Contents

Other manuals for KERN DBS 60-3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN DBS 60-3 and is the answer not in the manual?

KERN DBS 60-3 Specifications

General IconGeneral
BrandKERN
ModelDBS 60-3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals