Verification 4
Keysight 11713D/E Operating and Service Manual 41
Operator’s Check for Remote Operation
These procedure verify that the driver can be controlled remotely using GPIB, USB, and/or LAN.
1 Refer to Chapter 7, "Remote Interface Configurations", to connect the 11713D/E to your
computer through GPIB, USB, and/or LAN.
2 Once remote connection is available, send the following SCPI commands to the driver and note
the changes on front panel LEDs.
If the above checks are successful, the driver’s remote operation is working correctly. These
procedures do not check all of the driver’s program codes that can be executed. However, if the
driver work correctly from the front panel, there is a high probability that the driver will respond to
all the program codes.
Description Command Observation
Close switching paths from channel 1 to
channel 4 (bank1)
ROUTe:CLOSe (@101:104)
LEDs for
pushbuttons
1 to 4 light
ON
Open switching paths from channel 5 to
channel 8 (bank 1)
ROUTe:OPEn (@105:108)
LEDs for
pushbuttons
5 to 8 light
OFF
Query status on channel 2 (bank 1) ROUTe:CLOSe? (@102)
Return value “1” (LED for
pushbutton 2 light ON)
Query status on channel 7 (bank 1) ROUTe:CLOSe? (@107)
Return value “0” (LED for
pushbutton 7 light OFF)