6 Remote Operations
68 Keysight 11713D/E Operating and Service Manual
Commands relevant to 11713D/E
ROUTe:CLOSe
Closes the switching path(s). The first parameter specifies the Channels in the form of
‘bcc’ where b = bank number (0-n, range model dependent) and cc = channel number
(00-nn, range model dependent). For example: 103 or 104 or 108. For detailed
information, refer to Rules for using a channel or scan list section.
ROUTe:CLOSe <channel_list>
Examples
To close the channel number 106:
ROUTe:CLOSe (@106)
To query the closed list of channels:
ROUTe:CLOSe? <channel_list>
Related Commands
ROUTe:OPEn
ROUTe:CLOSe:ALL
ROUTe:OPEn:ALL
ROUTe:OPEn
Opens the switching path(s). The first parameter specifies the Channels in the form of
‘bcc’ where b = bank number (0-n, range model dependent) and cc = channel number
(00-nn, range model dependent). For example: 103 or 104 or 108. For detailed
information, refer to Rules for using a channel or scan list section.
ROUTe:OPEn <channel_list>
Query? Yes
Set? Yes
Set/Return Parameters
Numeric representation format: Integer (for
example, 8)
Minimum
101 (11713D)
101, 201 (11713E)
Maximum
110 (11713D)
110, 210 (11713E)
Units None
Query? Yes
Set? Yes
Set/Return Parameters
Numeric representation format: Integer (for
example, 8)
Minimum
101 (11713D)
101, 201 (11713E)

Table of Contents