Remote Operations 6
Keysight 11713D/E Operating and Service Manual 85
set the ESB (Event Summary Bit) of the Status Byte register. A “1” in the bit position
enables the corresponding event. All of the enabled events of the Standard Event Status
Event Register are logically ORed to cause the Event Summary Bit (ESB) of the Status
Byte Register to be set.
The query reads the Standard Event Status Enable register.
Table 6-1 Bit configuration of Standard Event Status Enable register
*ESE <Integer>
Examples
To set the bit in the enable register part of the standard event status register to 129:
*ESE 129
To query the enable register part of the standard event status register:
*ESE?
Related Commands
*ESR?
*SRE
*STB?
*ESR?
Reads and clears the Standard Event Status Event register. The bit configuration of this
register is the same as the Standard Event Status Enable register (refer to “*ESE”).
*ESR?
Bit Position 76543210
Bit Name PON n.u. CME EXE DDE QYE n.u. OPC
Bit Weight 128 32 16 8 4 1
n.u. - Not Used
PON- Power On
CME - Command Error
EXE - Execution Error
DDE - Device-dependent Error
QYE - Query Error
OPC - Operation Complete
Query? Yes
Set? Yes
Set/Return Parameters Integer (register value)
Minimum 0
Maximum 255
Units None