Getting Started 1
Keysight 53150A/151A/152A Operating Guide 49
Measuring Power (Channel 2 Only)
The Keysight 53150A/151A/152A Counter can measure signal power (in the same
frequency ranges as for frequency measurements) on Channel 2. The example
procedure for measuring power in the following diagram assumes that the
Counter has previously been set up to measure a 25 GHz input on Channel 2.
NOTE
When the power of a signal applied to the Channel 2 input exceeds the
maximum rated power for the instrument, the Counter displays HI.
CAUTION
The Channel 2 input path circuits contain sensitive GaAs semiconductors. To
prevent damage to these components, always adhere to standard ESD
(Electro-Static Discharge) prevention procedures, and ensure that the
maximum power specification for this channel (+27 dBm) is not exceeded.