EasyManua.ls Logo

Keysight Technologies B1505A

Keysight Technologies B1505A
128 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
109
side as shown in figure 2-8 (reside in main section).
For disconnecting the interlock cable, hold the metal part and then pull the
connector by turning it.
Step number 3:
Using a 16493L GNDU Cable, connect the GNDU on the B1505A to the GNDU1
Input on the N1259A.
Step number 4:
Using two 16494A Triax Cables, connect the Force and Sense connectors on the
lower B1510A HPSMU (SMU1) to the respective connectors on the HPSMU1
input of the N1259A.
Note:
HPSMU uses a pair of two triax cables. HCSMU cables looks similar as
HPSMU cables, but HCSMU cables are a pair of a coax cable and a triax cable.
It is a good practice for distinguishing these two cables.
Step number 5:
Using two 16494A Triax Cables, connect the upper B1510A HPSMU (SMU2) to
the HPSMU2 on the N1259A.
Step number 6:
Using a N1300A CMU Cable, connect the B1520A CMU to the respective connec-
tors of MF CMU Input (Hcur, Hpot, Lcur, Lpot) on the N1259A.
Note: Leave the green cable with a round terminal as it is.
Step number 7:
Using a 16493S HCSMU Cable, connect the Force and Sense connectors on the
B1512A HCSMU to the respective connectors of HCSMU1 Input on the N1259A.
Note:
HCSMU cables are a pair of a coax cable and a triax cable. In case of
HPSMU, it uses a pair of two triax cables. It is a good practice for distin-
guishing these two cables.
Step number 8:
Using a 16493T HV Triax Cable, connect the Force connector on the B1513A
HVSMU to the HVSMU1 input of the N1259A
Tips:
Make sure to screw in the connector firmly by hand to the end point where
you feel some resistance.
Step number 9:
Connect power cable to the fixture.

Other manuals for Keysight Technologies B1505A

Related product manuals