Measurement Parameters: Vce(sus) or V(BR)ceo Collector-Emitter sustaining
voltage or Collector-Emitter breakdown voltage
Iceo Collector-Emitter Cut-off current
Application Test name: Ic-Vceo_R
Application Test setup name: (My Favorite Setup -> Example_AT): Ic-Vceo_R or Ic-
Vceo_R(+R value)
Device used in the example: MJL4281AG
Connection inside the N1259A Test Fixture: Use the figure shown in this section
Application description:
Measures Collector current vs. Collector-Emitter voltage and extracts Collector-
Emitter breakdown voltage (BVceo) and cut-off current (Iceo) at open base condi-
tion. The voltage drop by a resistor0, 1k, 100 k) inserted between the collector
terminal and the collector SMU can be compensated. A reminder message for assur-
ing the resister value is popped up just before starting the measurement.
Measurement is aborted just after the detection of the drain breakdown.
For Vce(sus) or V(BR)ceo test, we have an option for inserting a 100 k resistor
between the collector and the collector SMU as shown in figure 3-13. Two configu-
rations are shown as an example for no resistor (0 selection) case and 100k
resistor case.
NOTE: If measurement is made with Opt.300 Module Selector, use the procedure
shown in Appendix 2.
Α.
Α.Α.
Α.
Connection inside the N1259A Test Fixture: 0
Use the following connection inside the N1259A test fixture
Open the N1259A test fixture cover, and connect the test leads shown in figure 3-2
by following the step numbers as shown in figure 3-11 and the following procedure
steps.
Note: The available colors of the leads are black and red only. The lead color used in
figure 3-11 is for reference only.
The numbers of the procedure steps correspond to the numbers on the drawing for
connecting the test leads.
3-1-4. Vce(sus) or V(BR)ceo, and Iceo-Vce measurement
Connection inside the N1259A Test Fixture

Other manuals for Keysight Technologies B1505A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B1505A and is the answer not in the manual?

Keysight Technologies B1505A Specifications

General IconGeneral
BrandKeysight Technologies
ModelB1505A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals