Chapter 2. Preparation for the Measurements
2-1. Before using B1505A
There are potentially hazardous voltages (3000 V for HVSMU, and 200 V
for HPSMU) present at the Force, Sense, and Guard terminals of Agilent
B1505A. To prevent electrical shock, the following safety precautions
must be observed during the use of B1505A.
Use a three-conductor AC power cable to connect the cabinet (if
used) and B1505A to an electrical ground (safety ground).
You must connect an interlock cable between B1505A and the test
fixture.
Confirm periodically that the interlock function is functional.
Do not modify the interlock circuit.
Do not use extension cables for connecting the DUT the outside of
the test fixture.
Before touching the connections on the Force, Guard, and Sense ter-
minals, turn the B1505A off and discharge any capacitors. If you do
not wish to turn the B1505A off, complete all of the following items,
regardless of the B1505A settings.
- Press the Stop key to turn the module output off.
- Confirm that the High Voltage indicator is not lit.
- Open the shielding box access door.
- Discharge any capacitors connected to an SMU.
WARNING

Other manuals for Keysight Technologies B1505A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B1505A and is the answer not in the manual?

Keysight Technologies B1505A Specifications

General IconGeneral
BrandKeysight Technologies
ModelB1505A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals