[PROCEDURE]
Open the N1259A test fixture cover, and connect the test leads as follows. The num-
bers on the drawing correspond to the procedure steps.
Step 1. Insert the device (MJL4281AG) into the socket on the N1259A.
Step 2. Connect the High Force of the Module Selector Output to the terminal 2
Force (Collector) on the Inline Package Socket.
Step 3. Connect the High Sense of the Module Selector Output to the terminal 2
Sense (Collector) on the Inline Package Socket.
Step 4. Connect the Low Force of the Module Selector Output to the terminal 3
Force (Emitter) on the Inline Package Socket.
Step 5. Connect the Low Sense of the Module Selector Output to the terminal 3
Sense (Emitter) on the Inline Package Socket.
Step 6. Connect the HPSMU2 Force to t to the terminal 1 Force (Base) on the Inline
Package Socket.
Step 7. Connect the HPSMU2 Sense to t to the terminal 1 Sense (Base) on the Inline
Package Socket.
Close the N1259A fixture cover.