Using the Solar Array Simulator 4
Series E4360 User’s Guide 81
Front Panel Keys
Lockout
You can lock the front panel keys to prevent unwanted control of the
instrument from the front panel. This is the most secure way of locking
the front panel keys because you need a password to unlock the front
panel. The lockout setting is saved in non-volatile memory so that the
front panel remains locked even after AC power is cycled.
Select System\Preferences\Lock
In the dialog box, enter the password
to unlock the front panel. Then select
Lock.
The menu to unlock the front panel
appears every time a key is pressed.
Enter the password to unlock the
front panel.
Not Available
If the password is lost, the SYSTem:PASSword:FPANel:RESet command
can reset the front panel lockout password to 0 (zero). It does not reset
the administration/calibration password. Refer to the Programmer’s
Reference Help file on your Keysight 14360 System Control Tools CD for
more information
NOTE
The SYSTem:COMMunicate:RLSTate RWLock command can also lock
and unlock the front panel. This command is co
the front panel lockout function. If you use this command to lock the
front panel, the front panel will be unlocked when AC power is cycled.
Keys
You can enable or disable the front panel key clicks.
Front Panel:
SCPI Command:
Select
Check Enable key clicks to enable key
clicks. Un-check to disable key clicks.
Not Available
You can configure the On/Off key to enable or disable ALL outputs.
Front Panel:
SCPI Command:
Select System\Preferences\Keys.
Check On/Off key affects all channels.
The ON/Off key will now be active on
ALL channels. The All indicator will be
on.
Not Available