Appendix A Specifications and Application Information
88 Series E4360 User’s Guide
Keysight E4360A SAS Mainframe
Supplemental Characteristics
Parameter
E4360A
≤ 1 ms from receipt of command to start of output change
Savable States
Memory locations:
Pre-stored state:
2 ( 0 and 1)
0
Protection Response Characteristics
INH input:
Fault on coupled outputs:
5 µs from receipt of inhibit to start of shutdown
< 10 µs from receipt of fault to start of shutdown
Digital Control Characteristics
Maximum voltage ratings:
Pins 1 and 2 as FLT output:
Pins 3 - 7 as digital/trigger outputs:
(pin 8 = common)
Pins 1 - 7 as digital/trigger inputs and pin
3 as INH input:
(pin 8 = common)
+16.5 VDC/
5 VDC between pins
(pin 8 is internally connected to chassis ground).
Maximum low-level output voltage = 0.5 V @ 4 mA
Maximum low-level sink current = 4 mA
Typical System-level leakage current = 1 mA @ 16.5 VDC
Maximum low-level output voltage = 0.5 V @ 4 mA;
1 V @ 50 mA; 1.75 V @ 100 mA
Maximum low-level sink current = 100 mA
Typical System-level leakage current = 0.8 mA @ 16.5 VDC
Maximum low-level input voltage = 0.8 V
Minimum System-level input voltage = 2 V
Typical low-level current = 2 mA @ 0 V (internal 2.2k pull-up)
Typical System-level leakage current = 0.12 mA @ 16.5 VDC
Interface Capabilities
GPIB:
LXI compliance:
USB 2.0:
10/100 LAN:
Built-in Web server:
SCPI - 1993, IEEE 488.2 compliant interface
Class C (only applies to units with LXI label on front panel)
Requires Keysight IO Library version M.01.01 or 14.0 and up
Requires Keysight IO Library version L.01.01 or 14.0 and up
Requires Internet Explorer 5+ or Netscape 6.2+
Environmental Conditions
Operating environment:
Temperature range:
Relative humidity:
Altitude:
Storage temperature:
Indoor use, installation category II (for AC input), pollution degree 2
0°C to 55°C (output current is derated above 40°C ambient temperature)
Up to 95%
Up to 2000 meters
-30°C to 70°C
Acoustic Noise Declaration
This statement is provided to comply with
the requirements of the German Sound
Emission Directive, from 18 January 1991.
Sound Pressure Lp <70 dB(A), At Operator Position,
Normal Operation, According to EN 27779 (Type Test).
Schalldruckpegel Lp <70 dB(A), Am Arbeitsplatz,
Normaler Betrieb, Nach EN 27779 (Typprüfung).