Performance Tests 2
Keysight E4416A/E4417A Power Meters Service Guide 39
Test Setup
Figure 2-4 Instrument Accuracy (Average Path) Test Setup
Procedure
Use the following procedure to perform the instrument accuracy (average path)
test:
1 Connect the equipment as shown in Figure 2-4.
2 Switch the power meter on.
3 Preset the instrument:
Press
then .
4 Set the display units to watts:
Press
, , .
Power Meter
CHANNEL A
Range Calibrator
Power sensor cable
NOTE
The procedure details the key presses required on the Keysight E4416A. For the
Keysight E4417A the equivalent key presses should be performed on both
channels.
Preset
Local
Confirm
Meas
Display
Units
W