Performance Tests 2
Keysight E4416A/E4417A Power Meters Service Guide 57
17 Switch to the power sensor's upper range:
– Press .
– Move to the Range parameter using the up/down arrow keys.
– Scroll through the Range values by pressing . Select a value of
UPPER.
– Press .
18 Verify that the display reads 0 dBm ±1 dB. Record the reading.
Table 2-6 Functional Test (Average Path) Results
Test Channel A Results Channel B Results
(Keysight E4417A Only)
Low Range, 30 dB pad _________________ _________________
Upper Range, 30 dB pad _________________ _________________
Low Range, no pad _________________ _________________
High Range, no pad _________________ _________________