EasyManuals Logo

Keysight Technologies E4417A User Manual

Keysight Technologies E4417A
163 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #52 background imageLoading...
Page #52 background image
Performance Tests 2
Keysight E4416A/E4417A Power Meters Service Guide 57
17 Switch to the power sensor's upper range:
Press .
Move to the Range parameter using the up/down arrow keys.
Scroll through the Range values by pressing . Select a value of
UPPER.
Press .
18 Verify that the display reads 0 dBm ±1 dB. Record the reading.
Channel
Change
Done
Table 2-6 Functional Test (Average Path) Results
Test Channel A Results Channel B Results
(Keysight E4417A Only)
Low Range, 30 dB pad _________________ _________________
Upper Range, 30 dB pad _________________ _________________
Low Range, no pad _________________ _________________
High Range, no pad _________________ _________________

Table of Contents

Other manuals for Keysight Technologies E4417A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E4417A and is the answer not in the manual?

Keysight Technologies E4417A Specifications

General IconGeneral
BrandKeysight Technologies
ModelE4417A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals