EasyManuals Logo

Keysight Technologies EDU36311A User Manual

Keysight Technologies EDU36311A
63 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #44 background imageLoading...
Page #44 background image
Using the Protection Function
Protection function
Each output has independent protection functions. A front panel status indicator will turn on when a protection
function is set. Protection functions are latching, which means that they must be cleared once they are set.
Of the following protection functions, only OV and OC are user-programmable.
OV: Over-voltage protection is a hardware OVP whose trip level is a programmable value. The OVP is always
enabled.
OC: Over-current protection is a programmable function that can be enabled or disabled. When enabled, the
output will be disabled when the output current reaches the current limit setting.
OT: Over-temperature protection monitors the temperature of each output and shuts down the output if any
temperature exceeds the maximum factory-defined limits.
Configuring protection
Protection functions are configured on the Source Settings window.
Press Source Settings to access the Source Settings window.
OV Protection
Over-voltage protection disables the output if the output voltage reaches the OVP level.
To set the over-voltage protection, enter an over-voltage value in the OV Protection field.
44
Keysight EDU36311A User's Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies EDU36311A and is the answer not in the manual?

Keysight Technologies EDU36311A Specifications

General IconGeneral
BrandKeysight Technologies
ModelEDU36311A
CategoryPower Supply
LanguageEnglish

Related product manuals