EasyManuals Logo

Keysight Technologies EL34243A User Manual

Keysight Technologies EL34243A
141 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #120 background imageLoading...
Page #120 background image
Scope marker view
Items Description
1 m1/m2 points Shows where the measurement markers intersect the selected waveform. Data values at the bottom of the dis-
play are referenced to the intersect locations of the markers. Calculations are based on the data points in
between the intersect locations.
2 Delta Indicates the delta or absolute difference between the markers in units (volts, amps, or watts) and in time
(seconds). The value in parenthesis is the frequency, which is the reciprocal of the time (1/time).
3 m2 Indicates the m2 marker value in volts, amps, or watts at the intersection point. Also indicates the distance in
time that the m2 marker is in relation to the present trigger position. If a marker is out of view, an arrow indic-
ates the direction of the marker .
4 m1 Indicates the m1 marker value in volts, amps, or watts at the intersection point. Also indicates the distance in
time of the m1 marker in relation to the present trigger position. If a marker is out of view, an arrow indicates
the direction of the marker .
5 Min Indicates the minimum data value (in volts, amps, or watts) between the marker locations of the selected wave-
form. Also indicates the distance in time of the minimum value in relation to the present trigger position.
6 Max Indicates the maximum data value (in volts, amps, or watts) between the marker locations of the selected
waveform. Also indicates the distance in time of the maximum value in relation to the present trigger position.
7 Avg. (if selected) Calculates the average data value (in volts, amps, or watts) between the marker locations of the selected wave-
form. Time indicates the time between markers over which the average value is calculated.
8 RMS (if selected) Calculates the rms value between the marker locations.
9 Vp-p (if selected) Calculates the difference between the maximum and minimum values. Time information is not valid for
calculated p-p values.
120
Keysight EL30000 Series User's Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies EL34243A and is the answer not in the manual?

Keysight Technologies EL34243A Specifications

General IconGeneral
BrandKeysight Technologies
ModelEL34243A
CategoryAccessories
LanguageEnglish

Related product manuals