Keysight N9927-90001 User’s Guide                                                            179
SA (Spectrum Analyzer) Mode
SA Mode Settings
—Press Default All to return all traces to their default settings: Trace 1 
(Clear/Write); all other traces (Blank).
Average Type
In SA Mode, there are four different processes in which Averaging is performed:
—Average Traces – Learn more in “Averaging” on page 195. 
— Detection Method Averaging – Learn more in “Detection Method” on 
page 188.
— Noise Marker Averaging - Learn more in “Noise Marker” on page 191.
— Video BW filtering (for non-Zero Span measurements) - Learn more in 
“Video Bandwidth (Video BW)” on page 166.
There are three types of mathematical averaging that can be performed – Log, 
Linear, or Voltage (see “Average Type” on page 179). Select ONE of these 
types and it is used for all of the above averaging processes. 
How to set Average Type
—Press Meas Setup 4 then Average Type [current setting]
— Then choose from the following:
—Auto
 - The FieldFox chooses the most appropriate type of averaging 
for the current settings. When Detection Method is set to Average, a 
Noise Marker is present, or a Channel measurement is active, then 
Power Average is ALWAYS selected. Otherwise, Log Average is 
selected.
—Log
 Averaging – Best for displaying Trace Averaging. LgAv is shown on 
the left side of the FieldFox screen when selected.
—Power
 (Linear) Averaging – Best for measuring true power levels. Used 
in Detection Average and Noise Marker Average. Mathematically, 
trace noise is 2.5 dB higher than when using Log Average. PAvg is 
shown on the left side of the FieldFox screen when selected.
—Voltage
 
Trace 4 data WILL be overwritten by the FieldFox when using the 
Independent Source Normalize feature (“Independent Source/Tracking 
Generator ” on page 163) or using Field Strength antenna or cable 
corrections (“Field Strength Measurements” on page 156).