SA (Spectrum Analyzer) Mode
How to Set Up EMF Settings (Requires EMF (Option 358), GPS (Option 307), and SA 
Mode (Option 233–Mixed Analyzers))
How to Set Up EMF Settings (Requires EMF (Option 358), GPS 
(Option 307), and SA Mode (Option 233–Mixed Analyzers))
EMF mode measures signals at the SA RF Input Port 2 connector. 
EMF mode requires the FieldFox to have SA mode (Option 233 on combination 
analyzers)), GPS (Option 307), and CPU2 installed. 
Recommended, but not required, preamplifier (Option 235) to improve receiver 
sensitivity.
With Option 358 (EMF), network designers can measure and verity the RF 
exposure compliance as required by each country. 
If SA mode is licensed for EMF Option 358, the primary measurement 
performed is Channel Power (CHP) which automatically switches amongst the 
3 TriAxial antenna axis (X/Y/Z), to provide a summation of "total power" 
detected, using unique Antenna supplied corrections for the individual axis, 
with results in FieldStrength units. This is accomplished within the existing SA 
Channel Power measurement.
Even without license EMF Option 358, the System Utilities menu enables one 
to Import Corrections and manually drive the Antenna to a particular axis (X, Y, 
or Z). The manual control thus allows any SA Receiver measurement to sweep 
that axis just like any other antenna transducer, however without license the 
imported corrections are not usable from the Apps. But with license the SA 
provides for easy use of the imported XYZ corrections, thus allowing 
presentation of correctly compensated FieldStrength measurements, for 1 axis 
at a time. 
This section reviews the steps in setting up an EMF measurement, using a USB 
antenna. If you are not familiar with this process or would like more 
information, refer to Chapter 14, “USB Antennas – (Full Capability Requires 
EMF Option 358, and SA mode (Option 233 Mixed Analyzers)).”
This section consists of:
If OTA or SA mode does not detect an EMF Option 358 license, then the 
EMF application and the USB antenna’s axes must be set up manually. This 
use model is not recommended or supported by Keysight. 
The USB antenna’s cable loss is not included, in the antenna’s correction 
factors. For an accurate EMF measurement, the cable loss values should 
be accounted for. Refer to “Editing Cable Correction Factors and Adding 
New Cable Factors” on page 215.
IMPORTANT! The USB triaxial antenna—or equivalent—is required for EMF 
measurements. Refer to Chapter 14, “USB Antennas – (Full Capability 
Requires EMF Option 358, and SA mode (Option 233 Mixed Analyzers)).”