62 Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide
2 Quick Reference
sample rate = number of samples / time of acquisition
For example, when storing 50 µs of data in 50,000 points of memory, the actual
sample rate is 1 GSa/s.
Likewise, when storing 50 ms of data in 50,000 points of memory, the actual
sample rate is 1 MSa/s.
The actual sample rate is displayed in the right-side information area.
The oscilloscope achieves the actual sample rate by throwing away (decimating)
unneeded samples.