38 Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide
2 Quick Reference
Vertical Controls
Vertical Knobs and Keys
Vertical Softkey Controls
The following figure shows the Channel 1 menu that appears after pressing the [1]
channel key.
NOTE
Keysight recommends always scaling the signal so that the entire waveform is contained
between the top and bottom of the display.
For proper operation of the 1000 X-Series oscilloscope, the channel inputs must not be
overdriven more than ±8 divisions. Exceeding this limit may result in signals that appear
incorrect and may increase crosstalk between the input channels.
NOTE
To minimize crosstalk between input channels, make sure the channel is not overdriven. Also,
connecting a probe or cable to a channel will reduce crosstalk.