Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide 123
Service and Utility Menus
Utility Service Menu
These adjustments are built into the instrument calibration and test harness. It
will take approximately ½ an hour for these adjustments to complete,
depending on the power meter/sensor used. Once this routine is started it
cannot be paused, only aborted. If aborted, it can only be started again from
the beginning.
For the last adjustment in this routine a power meter that can be controlled by
the instrument is required. This can be connected to the instrument either via
Sockets, VXI-11, or USB. However, the type of power meter/sensor used can
have a large effect on the amount of time the adjustment will take to complete.
Because of this, the recommended connection type is USB. When selecting a
power meter/sensor to use for this adjustment the following parameters need
to be considered:
Frequency Range: 10 MHz to 6 GHz
Power Level: -70 to +20 dBm
While there may be other sensors that meet the required parameters, the
recommended USB power sensors are shown in Table 5-2. When using one of
these recommended USB power sensors it will take approximately ½ an hour
for them all to complete.