332 Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide
Assembly Replacement
Before Replacing an Assembly
Before Replacing an Assembly
Before replacing any instrument assembly, perform the related “Pre Repair
Procedures” found in Chapter 16, “Pre and Post-Repair Procedures”. Doing so
will ensure that all required instrument information and calibration data will be
properly preserved once the assembly replacement is completed.
See the “Hardware Versus Firmware” in Chapter 19, “Instrument Firmware
and Operating System” for information on instrument firmware requirements
before replacing any electronic assembly.
Be sure to review the warning and caution statements described in Chapter 1,
“Overview” prior to replacing an assembly in your signal generator.
Many of the assemblies in this instrument are very susceptible to damage
from electrostatic discharge (ESD). Perform service procedures only at a
static-safe workstation and wear a grounding strap.