EasyManuals Logo

Keysight Technologies N5173B EXG Service Guide

Keysight Technologies N5173B EXG
451 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #340 background imageLoading...
Page #340 background image
340 Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide
Assembly Replacement
Inner Bottom Cover
Inner Bottom Cover
Tools Required
T-10 driver
Removal Procedure
Refer to Figure 15-3 for this procedure.
1. Disconnect the power cord.
2. Remove the outer-cover from the signal generator. Refer to “Outer Cover”
on page 336.
3. Place the signal generator flat with the front-panel facing you.
4. Using the T-10 driver, remove the eleven flat-top screws (1) from the
inner bottom cover (2).
5. Using the T-10 driver, remove the screw (3) from the rear panel.
6. Remove the inner bottom cover.
Replacement Procedure
1. Reverse the order of the removal procedures.
2. Recommended torque for all screws is 9 in–lbs.

Table of Contents

Other manuals for Keysight Technologies N5173B EXG

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N5173B EXG and is the answer not in the manual?

Keysight Technologies N5173B EXG Specifications

General IconGeneral
BrandKeysight Technologies
ModelN5173B EXG
CategoryPortable Generator
LanguageEnglish

Related product manuals