Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide 397
Assembly Replacement
A7 Micro-Deck (N5173B/83B only - Option 1EM)
Replacement Procedure
Refer to Figure 15-37 for this procedure.
1. Using the T-10 driver, remove the five screws (3) that secure the
replacement A7 Micro Deck to its packaging.
2. Carefully slide the replacement A7 Micro Deck out of its packaging.
3. Reverse the order of the removal procedure.
4. Torque all screws to 9 in–lbs.
5. Perform the post-repair procedures that pertain to this replacement
procedure. See Chapter 16, “Pre and Post-Repair Procedures.”