380 Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide
Assembly Replacement
Power Switch
Power Switch
Tools Required
— T-10 driver
Removal Procedure
Refer to Figure 15-29 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to “Outer Cover” on
page 336, “Inner Top Cover” on page 338, and “Inner Bottom Cover” on
page 340.
3. Remove the front-panel from the signal generator. Refer to “Front-Panel”
on page 342.
4. Remove the A6 Front Panel Interface Scan assembly from the signal
generator. Refer to “A6 Front Panel Interface Scan Assembly” on page 371.
5. Remove the LCD Display from the signal generator. Refer to “LCD Display”
on page 375.
6. Using the T-10 driver, remove the two screws (4) that secure the Power
Switch (2) to the front-panel frame.
7. Remove the Power Switch.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Recommended torque for all screws is 9 in–lbs.
3. Perform the post-repair procedures that pertain to this replacement