398 Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide
Assembly Replacement
A8 Floating BNC Bypass (N5173B/83B only)
A8 Floating BNC Bypass (N5173B/83B only)
Tools Required
— 5/8” hex-nut driver
— needle-nose pliers
Removal Procedure
Refer to Figure 15-38 for this procedure.
1. Disconnect the power cord.
2. Remove the inner cover and outer top cover from the signal generator.
Refer to “Outer Cover” on page 336, “Inner Top Cover” on page 338.
3. Disconnect the flexible cable from the Micro Deck at the location (1)
shown.
4. Using the 5/8” hex-nut driver, remove the nut and washer securing the A8
Floating BNC Bypass to the rear panel.
5. Remove the A8 Floating BNC Bypass.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Torque the nut to 21 in–lbs.