132 Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide
Troubleshooting
Unlocked Issues
phase lock loops to be unlocked. In this case both the reference and the
synthesizer unlock error messages would be present. So before
troubleshooting a synthesizer unlocked condition make sure that there is no
reference unlock error. If there is, resolve the reference unlock condition first.
For the CXG and EXG series instruments there are two possible synthesizer
unlock errors:
— 508 Synthesizer Unlocked; Sum Loop (over). The synthesizer is unlocked.
Service may be needed.
— 508 Synthesizer Unlocked; Sum Loop (under). The synthesizer is unlocked.
Service may be needed.
For the MXG series instruments there are five possible synthesizer unlock
errors:
— 508 Synthesizer Unlocked; FracN Loop. The synthesizer is unlocked.
Service may be needed.
— 508 Synthesizer Unlocked; Offset Loop (over). The synthesizer is unlocked.
Service may be needed.
— 508 Synthesizer Unlocked; Offset Loop (under). The synthesizer is unlocked.
Service may be needed.
— 508 Synthesizer Unlocked; Sum Loop (over). The synthesizer is unlocked.
Service may be needed.
— 508 Synthesizer Unlocked; Sum Loop (under). The synthesizer is unlocked.
Service may be needed.
As can be seen from the unlock errors listed, the phase lock loop that is
unlocked is clearly identified in each of the error messages. There are also
adjustments for each of the phase lock loops in the performance verification
and adjustment software.
Use the following procedure to troubleshoot a 508 Synthesizer Unlocked error:
1. Verify that there is no 512 Reference Unlocked or 515 Reference Missing
errors. If there are, resolve those issues first.
2. From the 508 Synthesizer Unlocked error message given, determine which
synthesizer phase lock loop is unlocked (FracN Loop, Offset Loop, or Sum
Loop).
3. Referring to Table 6-1, run the applicable synthesizer adjustments for the
phase lock loop that is unlocked.

Table of Contents

Other manuals for Keysight Technologies N5173B EXG

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N5173B EXG and is the answer not in the manual?

Keysight Technologies N5173B EXG Specifications

General IconGeneral
BrandKeysight Technologies
ModelN5173B EXG
CategoryPortable Generator
LanguageEnglish

Related product manuals