Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide 63
Instrument Information and Calibration Data
Calibration Data
A restore needs to be performed whenever an assembly with factory calibration
data saved on it is installed during a repair. This can also be useful if
adjustments have been performed, but not yet backed up, and for some reason
the data needs to be removed and have the instrument revert to the factory
calibration data.
The typical way to do this is to run a utility that has been provided in the
performance verification and adjustment software titled Calibration Array
Restore. However, the following remote command can also be used to perform
a restore:
:SERVice:C ALibrate:REST ore
When a calibration data restore is performed all overwrite files that exist will be
deleted and the factory data on the assemblies is copied into the working data
on the A5 CPU assembly.